Medical College of Wisconsin
CTSICores SearchResearch InformaticsREDCap

An in vitro comparison of the rake angles between K3 and ProFile endodontic file systems. J Endod 2005 Mar;31(3):180-2

Date

03/01/2005

Pubmed ID

15735464

DOI

10.1097/01.don.0000137638.17882.ce

Scopus ID

2-s2.0-14944371088 (requires institutional sign-in at Scopus site)   16 Citations

Abstract

The purpose of this study was to compare rake angles of the ProFile and K3 file systems. Twenty-five 40/0.06 taper files were obtained for each system. Five files from the same manufacturer were placed perpendicularly into a vial of Epoxicure Resin and left to set for 24 h. The set-ups were removed from the vials and each were sectioned 5 mm from the tip of the files and polished. A photomicrograph was taken of each file with 100x magnification. Five sets of ProFile and five sets of K3 files were processed in this manner. Images were captured digitally, and rake angles of each file were measured. Multivariate ANOVA found a significant difference (p < 0.001) among the three negative rake angles of the ProFile system compared with the K3 system.

Author List

Chow DY, Stover SE, Bahcall JK, Jaunberzins A, Toth JM

Author

Jeffrey M. Toth PhD Associate Dean for Research in the School of Dentistry department at Marquette University




MESH terms used to index this publication - Major topics in bold

Dental Instruments
Equipment Design
Multivariate Analysis
Root Canal Preparation