Dose perturbations due to contrast medium and air in mammosite treatment: an experimental and Monte Carlo study. Med Phys 2005 Jul;32(7):2279-87
Date
08/27/2005Pubmed ID
16121583DOI
10.1118/1.1943827Scopus ID
2-s2.0-22544453284 (requires institutional sign-in at Scopus site) 31 CitationsAbstract
In the management of early breast cancer, a partial breast irradiation technique called MammoSite (Proxima Therapeutic Inc., Alpharetta, GA) has been advocated in recent years. In MammoSite, a balloon implanted at the surgical cavity during tumor excision is filled with a radio-opaque solution, and radiation is delivered via a high dose rate brachytherapy source situated at the center of the balloon. Frequently air may be introduced during placement of the balloon and/or injection of the contrast solution into the balloon. The purpose of this work is to quantify as well as to understand dose perturbations due to the presence of a high-Z contrast medium and/or an air bubble with measurements and Monte Carlo calculations. In addition, the measured dose distribution is compared with that obtained from a commercial treatment planning system (Nucletron PLATO system). For a balloon diameter of 42 mm, the dose variation as a function of distance from the balloon surface is measured for various concentrations of a radio-opaque solution (in the range 5%-25% by volume) with a small volume parallel plate ion chamber and a micro-diode detector placed perpendicular to the balloon axis. Monte Carlo simulations are performed to provide a basic understanding of the interaction mechanism and the magnitude of dose perturbation at the interface near balloon surface. Our results show that the radio-opaque concentration produces dose perturbation up to 6%. The dose perturbation occurs mostly within the distances <1 mm from the balloon surface. The Plato system that does not include heterogeneity correction may be sufficient for dose planning at distances > or = 10 mm from the balloon surface for the iodine concentrations used in the MammoSite procedures. The dose enhancement effect near the balloon surface (<1 mm) due to the higher iodine concentration is not correctly predicted by the Plato system. The dose near the balloon surface may be increased by 0.5% per cm3 of air. Monte Carlo simulation suggests that the interface effect (enhanced dose near surface) is primarily due to Compton electrons of short range (<0.5 mm). For more accurate dosimetry in MammoSite delivery, the dose perturbation due to the presence of a radio-opaque contrast medium and air bubbles should be considered in a brachytherapy planning system.
Author List
Cheng CW, Mitra R, Li XA, Das IJMESH terms used to index this publication - Major topics in bold
ArtifactsBrachytherapy
Catheterization
Computer Simulation
Contrast Media
Female
Humans
Models, Biological
Models, Statistical
Monte Carlo Method
Radiation Injuries
Radiometry
Radiotherapy Dosage
Relative Biological Effectiveness
Scattering, Radiation