Operative Technique and Lessons Learned From Surgical Implantation of the NeuroPace Responsive Neurostimulation® System in 57 Consecutive Patients. Oper Neurosurg 2021 Jan 13;20(2):E98-E109
Date
10/20/2020Pubmed ID
33074294DOI
10.1093/ons/opaa300Scopus ID
2-s2.0-85099774213 (requires institutional sign-in at Scopus site) 31 CitationsAbstract
BACKGROUND: The Responsive Neurostimulation (RNS)® System (NeuroPace, Inc) is an implantable device designed to improve seizure control in patients with medically refractory focal epilepsy. Because it is relatively new, surgical pearls and operative techniques optimized from experience beyond a small case series have yet to be described.
OBJECTIVE: To provide a detailed description of our operative technique and surgical pearls learned from implantation of the RNS System in 57 patients at our institution. We describe our method for frame-based placement of amygdalo-hippocampal depth leads, open implantation of cortical strip leads, and open installation of the neurostimulator.
METHODS: We outline considerations for patient selection, preoperative planning, surgical positioning, incision planning, stereotactic depth lead implantation, cortical strip lead implantation, craniotomy for neurostimulator implantation, device testing, closure, and intraoperative imaging.
RESULTS: The median reduction in clinical seizure frequency was 60% (standard deviation 63.1) with 27% of patients achieving seizure freedom at last follow up (median 23.1 mo). No infections, intracerebral hemorrhages, or lead migrations were encountered. Two patients experienced lead fractures, and four lead exchanges have been performed.
CONCLUSION: The techniques set forth here will help with the safe and efficient implantation of these new devices.
Author List
Krucoff MO, Wozny TA, Lee AT, Rao VR, Chang EFAuthor
Max O. Krucoff BS, MD Associate Professor in the Neurosurgery department at Medical College of WisconsinMESH terms used to index this publication - Major topics in bold
Deep Brain StimulationDrug Resistant Epilepsy
Electrodes, Implanted
Epilepsies, Partial
Humans
Seizures