Intermediate-term results of a randomized clinical trial of the 350- versus the 500-mm2 Baerveldt implant. Ophthalmology 1994 Aug;101(8):1456-63; discussion 1463-4
Date
08/01/1994Pubmed ID
8058290DOI
10.1016/s0161-6420(94)31152-3Scopus ID
2-s2.0-0028000025 (requires institutional sign-in at Scopus site) 170 CitationsAbstract
BACKGROUND: The Baerveldt glaucoma implant is a large equatorial aqueous shunting device that is installed through a single-quadrant conjunctival incision. The intermediate-term results of a randomized study comparing the 350- and 500-mm2 Baerveldt implants are reported.
METHODS: Seventy-three patients with medically uncontrollable, nonneovascular glaucomas associated with aphakia, pseudophakia, or failed filters were enrolled in a randomized, prospective study comparing 350- and 500-mm2 Baerveldt implants. Surgical success was defined as 6 mmHg < or = final intraocular pressure < or = 21 mmHg without glaucoma reoperation or devastating complication.
RESULTS: Of patients with 350- and 500-mm2 implants, 93% and 88%, respectively, achieved surgical success (18-month life-table analysis, P = 0.93). The 500-mm2 implants afforded intraocular pressure control with significantly fewer medications (0.7 versus 1.3; P = 0.006). The postoperative visual acuities remained within one line of the preoperative visual acuities or improved in 62% and 66% of patients in the 350- and 500-mm2 groups, respectively (P = 0.93). Complication rates were statistically similar. The most frequent ones in the 350- and 500-mm2 groups, respectively, were serous choroidal effusion (16% and 32%), strabismus (16% and 19%), anterior uveitis (14% and 11%), and corneal or corneal graft edema (11% each).
CONCLUSION: The intermediate-term results of the 350- and 500-mm2 Baerveldt implants were statistically comparable with respect to surgical and visual outcomes, as well as complications, although the larger implant was associated with a higher rate of some complications. However, the 500-mm2 Baerveldt implant afforded intraocular pressure control with fewer medications than the 350-mm2 implant.
Author List
Lloyd MA, Baerveldt G, Fellenbaum PS, Sidoti PA, Minckler DS, Martone JF, LaBree L, Heuer DKAuthor
Dale Heuer MD Emeritus Professor in the Ophthalmology department at Medical College of WisconsinMESH terms used to index this publication - Major topics in bold
AdolescentAdult
Aged
Aged, 80 and over
Child
Drainage
Glaucoma
Humans
Middle Aged
Postoperative Complications
Prospective Studies
Prostheses and Implants
Treatment Outcome
Ultrasonography