Medical College of Wisconsin
CTSIResearch InformaticsREDCap

Noise analysis of three newborn infant isolettes. J Perinatol 1997;17(5):351-4

Date

11/28/1997

Pubmed ID

9373838

Scopus ID

2-s2.0-0031229377 (requires institutional sign-in at Scopus site)   7 Citations

Abstract

OBJECTIVE: The purpose of this study was to evaluate the noise of three newborn infant isolettes.

STUDY DESIGN: An observational, prospective study evaluated noise in three isolettes (Ohmeda, Air Shields, and Drager). The study measured interior isolette noise production during quiet noise situations and isolette noise attenuation of added low- and high-frequency noise. Noise was measured on the decibel A scale and in the 125 Hz and 1000 Hz bands.

RESULTS: During quiet conditions the Ohmeda and Drager isolettes had the least noise production on the basis of decibel A levels (p < 0.001). Low-frequency noise on the decibel A scale was attenuated the most by the Drager isolette (p < 0.001), although the overall difference may be clinically insignificant. The Ohmeda isolette attenuated high-frequency noise by 28.4 dB as measured on the decibel A scale, which was greater than values for the Drager (22.8 dB) and Air Shields (14 dB) isolettes (p < 0.001).

CONCLUSIONS: High-frequency noise attenuation by the Ohmeda and Drager isolettes is clinically and statistically greater than high-frequency noise attenuation by the Air Shields isolette.

Author List

Berens RJ, Weigle CG

Author

Richard J. Berens MD Professor in the Anesthesiology department at Medical College of Wisconsin




MESH terms used to index this publication - Major topics in bold

Evaluation Studies as Topic
Humans
Incubators, Infant
Infant, Newborn
Noise
Prospective Studies