A Holter-type, microprocessor-based, rehabilitation instrument for acquisition and storage of plantar pressure data. J Rehabil Res Dev 1997 Apr;34(2):187-94
Date
04/01/1997Pubmed ID
9108345Scopus ID
2-s2.0-0031127592 (requires institutional sign-in at Scopus site) 42 CitationsAbstract
A Holter-type, microprocessor-based, portable, in-shoe, plantar pressure data acquisition system has been developed. The system allows continuous recording of pressure data between the sole of the foot and the shoe during the performance of daily living activities. Fourteen conductive polymer sensors acquire the plantar pressure history, which is then stored in the system memory. Pressures are sampled at a rate of 40 Hz from each of the 14 sensors for up to 8 hrs. The extended recording and processing capacity of the system developed in this study allows quantitative analysis of cumulative plantar pressure and temporal gait data necessary for characterization of event-related alterations in plantar pressures. The alterations that could be examined with the system include rehabilitative, therapeutic, surgical, and nonsurgical treatment. The system is fully portable and does not disrupt the natural gait pattern of the subject during ambulation. Peak plantar pressures, pressure-time integrals, and contact durations are determined for each of the insole sensors.
Author List
Abu-Faraj ZO, Harris GF, Abler JH, Wertsch JJAuthors
Gerald F. Harris Emeritus Professor in the Orthopaedic Surgery department at Medical College of WisconsinGerald Harris PhD Director in the Orthopaedic Research Engineering Center (OREC) department at Marquette University
MESH terms used to index this publication - Major topics in bold
Equipment DesignFoot
Humans
Microcomputers
Pressure
Shoes
Signal Processing, Computer-Assisted