Portable Detection of Copper Ion Using Personal Glucose Meter. Sensors (Basel) 2024 Oct 31;24(21)

Date

11/13/2024

Pubmed ID

39517898

Pubmed Central ID

PMC11548145

DOI

10.3390/s24217002

Scopus ID

2-s2.0-85208578596 (requires institutional sign-in at Scopus site)   1 Citation

Abstract

A simple and sensitive method for Cu2+ detection was developed using the Cu+-catalyzed alkyne-azide cycloaddition reaction, Fe3O4 magnetic nanoparticles (MNPs) as the reaction platform, and a portable blood glucose meter (PGM) as the detection method. Gold nanoparticles (AuNPs) were labeled with glucose oxidase (GOx) and alkyne-functionalized, terminally thiolated ssDNA (C2). In the presence of Cu2+ and ascorbate, the functionalized AuNPs were captured onto MNPs modified with azide-functionalized ssDNA (C1) via the Cu+-catalyzed alkyne-azide cycloaddition reaction. The GOx on the AuNPs' surface oxidized glucose (Glu) into gluconic acid and H2O2, reducing the Glu content in the reaction solution, which was quantitatively detected by the PGM. Under optimal conditions, the PGM response of the system showed a good linear relationship with the logarithm of Cu2+ concentration in the range of 0.05 to 10.00 μmol/L, with a detection limit of 0.03 μmol/L (3σ). In actual tap water samples, the spiked recovery rate of Cu2+ was between 92.30% and 113.33%, and the relative standard deviation was between 0.14% and 0.34%, meeting the detection requirements for Cu2+ in real water samples.

Author List

Du B, Chen T, Huang A, Chen H, Liu W

Author

Wei Liu PhD Professor in the Pharmacology and Toxicology department at Medical College of Wisconsin




MESH terms used to index this publication - Major topics in bold

Alkynes
Azides
Biosensing Techniques
Blood Glucose
Blood Glucose Self-Monitoring
Copper
DNA, Single-Stranded
Glucose
Glucose Oxidase
Gold
Humans
Hydrogen Peroxide
Limit of Detection
Magnetite Nanoparticles
Metal Nanoparticles