Accuracy and precision of intraocular injection volume. Am J Ophthalmol 2002 Apr;133(4):564-6
Date
04/05/2002Pubmed ID
11931796DOI
10.1016/s0002-9394(01)01408-8Scopus ID
2-s2.0-0036207045 (requires institutional sign-in at Scopus site) 21 CitationsAbstract
PURPOSE: To assess the accuracy and precision of methods in clinical use for intraocular drug injection.
DESIGN: Experimental laboratory investigation.
METHODS: A total of 800 injections were performed using 20 different 1-ml syringes at intended volumes of 0.10 ml and 0.05 ml. The volume of each injection was calculated from mass of water delivered onto an analytic balance.
RESULTS: The mean measured volume delivered was less than the intended volume for every syringe (P <.001 for each syringe at each volume). A total of 7% and 22% of injections differed by 20% or more from the intended volumes of 0.10 ml and 0.05 ml, respectively. Overdelivery of volume was uncommon.
CONCLUSION: Clinically relevant materials and methods used for intraocular injection of drugs systematically underdelivered the intended volume.