Multi-Frequency Electrocochleography Results in Fewer Drop Alarms During Cochlear Implant Insertion. Laryngoscope 2026 Oct;136(10):4474-4481
Date
05/05/2026Pubmed ID
42084107Pubmed Central ID
PMC13569684DOI
10.1002/lary.70592Scopus ID
2-s2.0-105037881892 (requires institutional sign-in at Scopus site)Abstract
OBJECTIVE: To evaluate intracochlear electrocochleography (ECochG) amplitude parameters during cochlear implantation (CI) using a novel multi-frequency ECochG algorithm.
METHODS: A multi-institutional, prospective cohort study was performed at 18 high-volume CI centers. The inclusion criteria were adults with sensorineural hearing loss and audiometric thresholds of ≤ 90 dB hearing level at 500 Hz undergoing CI with Advanced Bionics (Valencia, CA) Ultra 3D devices between 2024 and 2025. ECochG recordings were performed with simultaneous multi-frequency stimulation of four frequencies between 125 and 4000 Hz during cochlear implant insertion. Concurrent multi-frequency recording allowed extraction of amplitude and phase of each frequency individually. Post hoc analysis was performed to determine the difference in the number of drop alarms between single- and multi-frequency ECochG. An ECochG amplitude drop of 6 dB was defined as a drop alarm. Insertion track patterns were compared between single- and multi-frequency ECochG.
RESULTS: One hundred ninety-five ears were included. Mean number of drop alarms for the single-frequency algorithm was 1.72 (95% CI: 1.52, 1.92; median 1) compared to 0.42 (95% CI: 0.31, 0.53; median 0) for multi-frequency; p < 0.001. The number of Type C patterns (rise in amplitude during insertion followed by a drop) decreased with the multi-frequency ECochG algorithm compared to the single-frequency ECochG algorithm. The number of Type D patterns (no-response) decreased, indicating that multi-frequency ECochG generated more responses across the cochlea than single-frequency ECochG.
CONCLUSIONS: A novel multi-frequency ECochG algorithm during CI is associated with fewer drop alarms and altered insertion track patterns, which may provide a more accurate assessment of the cochlear microenvironment.
Author List
Espahbodi M, Syed A, Tu A, Seidman MD, Danner CJ, Allen KP, Bartels LJ, Ahsan SF, Voelker CCJ, Deep NL, Hunter JB, Chiffer RC, Michaelides EM, Elrakhawy M, Miller MW, Lin HW, Djalilian HR, Vivas EX, Angeli SI, Hoa M, Kim HJ, Widick MH, Stidham KR, Patel NS, Gurgel RK, Doerfer KW, Koka K, Harris MSAuthors
Karl W. Doerfer MD Assistant Professor in the Otolaryngology and Communication Sciences department at Medical College of WisconsinMichael S. Harris MD Chief, Associate Professor in the Otolaryngology and Communication Sciences department at Medical College of Wisconsin
MESH terms used to index this publication - Major topics in bold
AlgorithmsAudiometry, Evoked Response
Clinical Alarms
Cochlear Implantation
Cochlear Implants
Female
Hearing Loss, Sensorineural
Humans
Prospective Studies









