Medical College of Wisconsin
CTSIResearch InformaticsREDCap

Investigation of force decay in aesthetic, fibre-reinforced composite orthodontic archwires. Eur J Orthod 2015 Feb;37(1):43-8

Date

08/03/2014

Pubmed ID

25086028

DOI

10.1093/ejo/cju003

Scopus ID

2-s2.0-84922475666 (requires institutional sign-in at Scopus site)   10 Citations

Abstract

BACKGROUND/OBJECTIVES: Because polymer-based materials typically exhibit viscoelastic properties, the objective was to determine if commercially available, aesthetic, fibre-reinforced composite archwires maintain continuous forces without undergoing force decay when deflected continuously.

MATERIALS/METHODS: Quasi force decay was evaluated by comparing three-point bending profiles of nickel-titanium (NiTi) and fibre-reinforced composite archwires (BioMers) prior to and after 30 days of continuous deflection of either 1 or 2mm. Paired t-tests or non-parametric signed rank tests were used to statistically compare pre- and post-deflection bending forces. A control group consisting of wires not subject to the 30-day constant deflection was tested to check whether the initial testing altered the second three-point bend test.

RESULTS: Significant (P < 0.01) differences in the pre- and post-deflection deactivation force delivery were most evident in the composite 2mm deflection group and all of the NiTi groups. The composite 2mm deflection group failed to deliver consistent forces as the majority of the wires experienced crazing during the 30-day deflection period. The decrease in force delivery in the NiTi groups may be attributed to the small standard deviations.

CONCLUSIONS: The composite 1mm deflection group demonstrated that fibre-reinforced composite archwires are able to deliver a consistent force after 30 days of deflection. However, the clinical applicability of these fibre-reinforced composite archwires may be limited as they are unable to sustain deflections of 2mm without experiencing crazing and loss of force delivery.

LIMITATIONS: Clinical efficacy of the aesthetic, fibre-reinforced composite orthodontic archwires remains to be observed.

Author List

Spendlove J, Berzins DW, Pruszynski JE, Ballard RW

Author

David Berzins BS,PhD Graduate Program Director for Dental Biomaterials in the General Dental Sciences/Dental Biomaterials department at Marquette University




MESH terms used to index this publication - Major topics in bold

Composite Resins
Dental Alloys
Dental Stress Analysis
Elasticity
Equipment Design
Esthetics, Dental
Humans
Materials Testing
Mechanical Phenomena
Nickel
Orthodontic Wires
Titanium