Comparison of the artifacts caused by metallic implants in breast MRI using dual-echo dixon versus conventional fat-suppression techniques. AJR Am J Roentgenol 2014 Sep;203(3):W307-14
Date
08/26/2014Pubmed ID
25148189DOI
10.2214/AJR.13.10791Scopus ID
2-s2.0-84907878909 (requires institutional sign-in at Scopus site) 20 CitationsAbstract
OBJECTIVE: The purpose of this article is to evaluate and compare the artifacts caused by metal implants in breast MR images acquired with dual-echo Dixon and two conventional fat-suppression techniques.
SUBJECTS AND METHODS: Two types of biopsy markers were embedded into a uniform fat-water emulsion. T1-weighted gradient-echo images were acquired on a clinical 3-T MRI scanner with three different fat-suppression techniques-conventional or quick fat saturation, spectrally selective adiabatic inversion recovery (SPAIR), and dual-echo Dixon-and the 3D volumes of artifacts were measured. Among the subjects of a clinical breast MRI study using the same scanner, five patients were found to have one or more metal implants. The artifacts in Dixon and SPAIR fat-suppressed images were evaluated by three radiologists, and the results were compared with those of the phantom study.
RESULTS: In the phantom study, the artifacts appeared as interleaved bright and dark rings on SPAIR and quick-fat-saturation images, whereas they appeared as dark regions with a thin bright rim on Dixon images. The artifacts imaged with the Dixon technique had the smallest total volume. However, the reviewers found larger artifact diameters on patient images using the Dixon sequence because only the central region was recognized as an artifact on the SPAIR images.
CONCLUSION: Metal implants introduce artifacts of different types and sizes, according to the different fat-suppression techniques used. The dual-echo Dixon technique produces a larger central void, allowing the implant to be easily identified, but presents a smaller overall artifact volume by obscuring less area in the image, according to a quantitative phantom study.
Author List
Le Y, Kipfer HD, Majidi SS, Holz S, Lin CAuthor
Shadie S. Majidi MD Chief, Professor in the Radiology department at Medical College of WisconsinMESH terms used to index this publication - Major topics in bold
Adipose TissueArtifacts
Breast Implants
Breast Neoplasms
Female
Humans
Image Enhancement
Image Interpretation, Computer-Assisted
Magnetic Resonance Imaging
Metals
Phantoms, Imaging
Reproducibility of Results
Sensitivity and Specificity
Subtraction Technique