Clinical evaluation of the Oculab Tono-Pen. Am J Ophthalmol 1987 Aug 15;104(2):168-73
Date
08/15/1987Pubmed ID
3618714DOI
10.1016/0002-9394(87)90010-9Scopus ID
2-s2.0-0023267991 (requires institutional sign-in at Scopus site) 111 CitationsAbstract
We performed a clinical evaluation on a prototype version of the Tono-Pen hand-held tonometer. A masked, randomized design was used to compare the instrument with the Goldmann tonometer. We tested 270 eyes. For pressures of 6 to 24 mm Hg, the Tono-Pen tonometer measured, on average, 1.7 mm Hg higher than the Goldmann tonometer. This difference was statistically significant (P less than .0001). In the range above 24 mm Hg, no statistically significant difference was found in the pressures measured with the two instruments. Measurements made with the Tono-Pen were sufficiently close to those made with the Goldmann tonometer to be considered clinically accurate.
Author List
Minckler DS, Baerveldt G, Heuer DK, Quillen-Thomas B, Walonker AF, Weiner JAuthor
Dale Heuer MD Emeritus Professor in the Ophthalmology department at Medical College of WisconsinMESH terms used to index this publication - Major topics in bold
Evaluation Studies as TopicHumans
Intraocular Pressure
Ophthalmology