Four-dimensional spectral-spatial imaging using projection reconstruction. IEEE Trans Med Imaging 1995;14(2):374-83
Date
01/01/1995Pubmed ID
18215840DOI
10.1109/42.387718Scopus ID
2-s2.0-0029325043 (requires institutional sign-in at Scopus site) 9 CitationsAbstract
An n-dimensional (n-D) filtered backprojection image reconstruction algorithm has been developed and used in the reconstruction of 4-D spectral-spatial magnetic resonance imaging (MRI) data. The algorithm uses n-1 successive stages of 2-D filtered backprojection to reconstruct an n-D image. This approach results in a reduction in computational time on the order of N(n-2) relative to the single-stage technique, where N(n) is the number of elements in an n-D image. The authors describe implementation of the algorithm, including digital filtering and sampling requirements. Images obtained from simulated data are presented to illustrate the accuracy and potential utility of the technique.