A semi-empirical model for the therapeutic range shift estimation caused by inhomogeneities in proton beam therapy. J Appl Clin Med Phys 2012 Mar 08;13(2):3631

Date

03/10/2012

Pubmed ID

22402381

Pubmed Central ID

PMC5716409

DOI

10.1120/jacmp.v13i2.3631

Scopus ID

2-s2.0-84860747907 (requires institutional sign-in at Scopus site)   13 Citations

Abstract

The purpose of this study was to devise a simple semi-empirical model to estimate the range shift in clinical practices with high-Z inhomogeneity in proton beam. A semi-empirical model utilizing the logarithmic dependence on Z in stopping power from Bohr's classical approach has been developed to calculate the range shift due to the presence of inhomogeneity. Range shift from metallic plates of atomic number Z of various thicknesses were measured in water using a parallel plate ionization chamber and calculated with the FLUKA Monte Carlo code. The proton range shifts for bone and polymethyl methacrylate (PMMA) were estimated using the semi-empirical model and compared with Monte Carlo calculation. The semi-empirical equation to determine range shift and water equivalent thickness is presented. The model predicts a shift of the therapeutic range to within 2.5% accuracy for initial proton energies of 50 to 250 MeV and atomic numbers from 3.3 (effective Z for water) to 82. This equation is independent of beam energy, and thus provides range shift from high-Z materials without the knowledge of proton energy. The proposed method of calculating the therapeutic range shift accurately requires only knowledge of the effective or actual atomic number of the inhomogeneity and the thickness of the inhomogeneity along the beam direction. The model generalizes the range shift calculation for any material based on its effective atomic number, and permits reliable prediction of the range shift for material combinations where no data is currently available. The proposed model can be readily implemented in routine clinical practice for proton range shift estimation and quality assurance on the treatment planning.

Author List

Moskvin V, Cheng CW, Fanelli L, Zhao L, Das IJ

Author

Li Zhao PhD Associate Professor in the Radiation Oncology department at Medical College of Wisconsin




MESH terms used to index this publication - Major topics in bold

Algorithms
Bone Neoplasms
Computer Simulation
Humans
Models, Theoretical
Monte Carlo Method
Phantoms, Imaging
Proton Therapy
Radiotherapy Planning, Computer-Assisted
Radiotherapy, High-Energy