Incomplete chest wall decompression: a clinical evaluation of CPR performance by EMS personnel and assessment of alternative manual chest compression-decompression techniques. Resuscitation 2005 Mar;64(3):353-62

Date

03/01/2005

Pubmed ID

15733766

DOI

10.1016/j.resuscitation.2004.10.007

Scopus ID

2-s2.0-20044364340 (requires institutional sign-in at Scopus site)   191 Citations

Abstract

BACKGROUND: Complete chest wall recoil improves hemodynamics during cardiopulmonary resuscitation (CPR) by generating relatively negative intrathoracic pressure and thus draws venous blood back to the heart, providing cardiac preload prior to the next chest compression phase.

OBJECTIVE: Phase I was an observational case series to evaluate the quality of chest wall recoil during CPR performed by emergency medical services (EMS) personnel on patients with an out-of-hospital cardiac arrest. Phase II was designed to assess the quality of CPR delivered by EMS personnel using an electronic test manikin. The goal was to determine if a change in CPR technique or hand position would improve complete chest wall recoil, while maintaining adequate duty cycle, compression depth, and correct hand position placement. Standard manual CPR and three alternative manual CPR approaches were assessed.

METHODS AND RESULTS: Phase I--The clinical observational study was performed by an independent observer noting incomplete chest wall decompression and correlating that observation with electronically measured airway pressures during CPR in adult patients with out-of-hospital cardiac arrest. Rescuers were observed to maintain some residual and continuous pressure on the chest wall during the decompression phase of CPR, preventing full chest wall recoil, at some time during resuscitative efforts in 6 (46%) of 13 consecutive adults (average +/- S.D. age 63 +/-5.8 years). Airway pressures were consistently positive during the decompression phase (>0 mmHg) during those observations. Phase II: This randomized prospective trial was performed on an electronic test manikin. Thirty EMS providers (14 EMT-Basics, 5 EMT-Intermediates, and 11 EMT-Paramedics), with an average age +/- S.D. of 32 +/- 8 years and 6.5 +/- 4.2 years of EMS experience, performed 3 min of CPR on a Laerdal Skill Reporter CPR manikin using the Standard Hand Position followed by 3 min of CPR (in random order) using three alternative CPR techniques: (1) Two-Finger Fulcrum Technique--lifting the heel of the hand slightly but completely off the chest during the decompression phase of CPR using the thumb and little finger as a fulcrum; (2) Five-Finger Fulcrum Technique--lifting the heel of the hand slightly but completely off the chest during the decompression phase of CPR using all five fingers as a fulcrum; and (3) Hands-Off Technique--lifting the heel and all fingers of the hand slightly but completely off the chest during the decompression phase of CPR. These EMS personnel did not know the purpose of the studies prior to or during this investigation. Adequate compression depth was poor for all hand positions tested and ranged only from 29.9 to 48.5% of all compressions. When compared with the Standard Hand Position, the Hands-Off Technique decreased mean compression duty cycle from 46.9 +/- 6.4% to 33.3 +/- 4.6%, (P < 0.0001) but achieved the highest rate of complete chest wall recoil (95.0% versus 16.3%, P < 0.0001) and was 129 times more likely to provide complete chest wall recoil (OR: 129.0; CI: 43.4-382.0). There were no significant differences in accuracy of hand placement, depth of compression, or reported increase in fatigue or discomfort with its use compared with the Standard Hand Position.

CONCLUSIONS: Incomplete chest wall decompression was observed at some time during resuscitative efforts in 6 (46%) of 13 consecutive adult out-of-hospital cardiac arrests. The Hands-Off Technique decreased compression duty cycle but was 129 times more likely to provide complete chest wall recoil (OR: 129.0; CI: 43.4-382.0) compared to the Standard Hand Position without differences in accuracy of hand placement, depth of compression, or reported increase in fatigue or discomfort with its use. All forms of manual CPR tested (including the Standard Hand Position) in professional EMS rescuers using a recording manikin produced an inadequate depth of compression more than half the time. Thes

Author List

Aufderheide TP, Pirrallo RG, Yannopoulos D, Klein JP, von Briesen C, Sparks CW, Deja KA, Conrad CJ, Kitscha DJ, Provo TA, Lurie KG

Author

Tom P. Aufderheide MD Professor in the Emergency Medicine department at Medical College of Wisconsin




MESH terms used to index this publication - Major topics in bold

Adult
Aged
Aged, 80 and over
Cardiopulmonary Resuscitation
Emergency Medical Services
Female
Heart Arrest
Humans
Male
Manikins
Middle Aged
Pressure
Thoracic Wall