Medical College of Wisconsin
CTSICores SearchResearch InformaticsREDCap

Technical Note: Development and performance of a software tool for quality assurance of online replanning with a conventional Linac or MR-Linac. Med Phys 2016 Apr;43(4):1713

Date

04/03/2016

Pubmed ID

27036569

DOI

10.1118/1.4943795

Scopus ID

2-s2.0-84962249160 (requires institutional sign-in at Scopus site)   30 Citations

Abstract

PURPOSE: To develop an integrated quality assurance (QA) software tool for online replanning capable of efficiently and automatically checking radiation treatment (RT) planning parameters and gross plan quality, verifying treatment plan data transfer from treatment planning system (TPS) to record and verify (R&V) system, performing a secondary monitor unit (MU) calculation with or without a presence of a magnetic field from MR-Linac, and validating the delivery record consistency with the plan.

METHODS: The software tool, named ArtQA, was developed to obtain and compare plan and treatment parameters from both the TPS and the R&V system database. The TPS data are accessed via direct file reading and the R&V data are retrieved via open database connectivity and structured query language. Plan quality is evaluated with both the logical consistency of planning parameters and the achieved dose-volume histograms. Beams in between the TPS and R&V system are matched based on geometry configurations. To consider the effect of a 1.5 T transverse magnetic field from MR-Linac in the secondary MU calculation, a method based on modified Clarkson integration algorithm was developed and tested for a series of clinical situations.

RESULTS: ArtQA has been used in their clinic and can quickly detect inconsistencies and deviations in the entire RT planning process. With the use of the ArtQA tool, the efficiency for plan check including plan quality, data transfer, and delivery check can be improved by at least 60%. The newly developed independent MU calculation tool for MR-Linac reduces the difference between the plan and calculated MUs by 10%.

CONCLUSIONS: The software tool ArtQA can be used to perform a comprehensive QA check from planning to delivery with conventional Linac or MR-Linac and is an essential tool for online replanning where the QA check needs to be performed rapidly.

Author List

Chen GP, Ahunbay E, Li XA

Authors

Ergun Ahunbay PhD Professor in the Radiation Oncology department at Medical College of Wisconsin
Guang-Pei Chen PhD Associate Professor in the Radiation Oncology department at Medical College of Wisconsin




MESH terms used to index this publication - Major topics in bold

Humans
Magnetic Resonance Imaging
Particle Accelerators
Quality Control
Radiometry
Radiotherapy Planning, Computer-Assisted
Software