An online replanning method using warm start optimization and aperture morphing for flattening-filter-free beams. Med Phys 2016 Aug;43(8):4575

Date

08/05/2016

Pubmed ID

27487874

DOI

10.1118/1.4955439

Scopus ID

2-s2.0-84979052955 (requires institutional sign-in at Scopus site)   12 Citations

Abstract

PURPOSE: In a situation where a couch shift for patient positioning is not preferred or prohibited (e.g., MR-linac), segment aperture morphing (SAM) can address target dislocation and deformation. For IMRT/VMAT with flattening-filter-free (FFF) beams, however, SAM method would lead to an adverse translational dose effect due to the beam unflattening. Here the authors propose a new two-step process to address both the translational effect of FFF beams and the target deformation.

METHODS: The replanning method consists of an offline and an online step. The offline step is to create a series of preshifted-plans (PSPs) obtained by a so-called "warm start" optimization (starting optimization from the original plan, rather than from scratch) at a series of isocenter shifts. The PSPs all have the same number of segments with very similar shapes, since the warm start optimization only adjusts the MLC positions instead of regenerating them. In the online step, a new plan is obtained by picking the closest PSP or linearly interpolating the MLC positions and the monitor units of the closest PSPs for the shift determined from the image of the day. This two-step process is completely automated and almost instantaneous (no optimization or dose calculation needed). The previously developed SAM algorithm is then applied for daily deformation. The authors tested the method on sample prostate and pancreas cases.

RESULTS: The two-step interpolation method can account for the adverse dose effects from FFF beams, while SAM corrects for the target deformation. Plan interpolation method is effective in diminishing the unflat beam effect and may allow reducing the required number of PSPs. The whole process takes the same time as the previously reported SAM process (5-10 min).

CONCLUSIONS: The new two-step method plus SAM can address both the translation effects of FFF beams and target deformation, and can be executed in full automation except the delineation of target contour required by the SAM process.

Author List

Ahunbay EE, Ates O, Li XA

Author

Ergun Ahunbay PhD Professor in the Radiation Oncology department at Medical College of Wisconsin




MESH terms used to index this publication - Major topics in bold

Algorithms
Automation
Humans
Male
Organs at Risk
Pancreatic Neoplasms
Prostatic Neoplasms
Radiotherapy Dosage
Radiotherapy Planning, Computer-Assisted
Radiotherapy, Intensity-Modulated