On the improvement of CBCT image quality for soft tissue-based SRS localization. J Appl Clin Med Phys 2018 Nov;19(6):177-184
Date
10/09/2018Pubmed ID
30294838Pubmed Central ID
PMC6236842DOI
10.1002/acm2.12470Scopus ID
2-s2.0-85054596767 (requires institutional sign-in at Scopus site) 7 CitationsAbstract
PURPOSE: We explore the optimal cone-beam CT (CBCT) acquisition parameters to improve CBCT image quality to enhance intracranial stereotactic radiosurgery (SRS) localization and also assess the imaging dose levels associated with each imaging protocol.
METHODS: Twenty-six CBCT acquisition protocols were generated on an Edge® linear accelerator (Varian Medical Systems, Palo Alto, CA) with different x-ray tube current and potential settings, gantry rotation trajectories, and gantry rotation speeds. To assess image quality, images of the Catphan 504 phantom were analyzed to evaluate the following image quality metrics: uniformity, HU constancy, spatial resolution, low contrast detection, noise level, and contrast-to-noise ratio (CNR). To evaluate the imaging dose for each protocol, the cone-beam dose index (CBDI) was measured. To validate the phantom results, further analysis was performed with an anthropomorphic head phantom as well as image data acquired for a clinical SRS patient.
RESULTS: The Catphan data indicates that adjusting acquisition parameters had direct effects on the image noise level, low contrast detection, and CNR, but had minimal effects on uniformity, HU constancy, and spatial resolution. The noise level was reduced from 34.5 ± 0.3 to 18.5 ± 0.2 HU with a four-fold reduction in gantry speed, and to 18.7 ± 0.2 HU with a four-fold increase in tube current. Overall, the noise level was found to be proportional to inverse square root of imaging dose, and imaging dose was proportional to the product of total tube current-time product and the cube of the x-ray potential. Analysis of the anthropomorphic head phantom data and clinical SRS imaging data also indicates that noise is reduced with imaging dose increase.
CONCLUSIONS: Our results indicate that optimization of the imaging protocol, and thereby an increase in the imaging dose, is warranted for improved soft-tissue visualization for intracranial SRS.
Author List
Mao W, Gardner SJ, Snyder KC, Wen NW, Zhong H, Li H, Jackson P, Shah M, Chetty IJAuthor
Hualiang Zhong PhD Associate Professor in the Radiation Oncology department at Medical College of WisconsinMESH terms used to index this publication - Major topics in bold
Bone and BonesCone-Beam Computed Tomography
Head
Humans
Image Processing, Computer-Assisted
Organs at Risk
Phantoms, Imaging
Prognosis
Radiosurgery
Radiotherapy Dosage
Radiotherapy Planning, Computer-Assisted
Radiotherapy, Intensity-Modulated
Soft Tissue Neoplasms