Correlation of Thromboelastography with Apparent Rivaroxaban Concentration: Has Point-of-Care Testing Improved? Anesthesiology 2020 Feb;132(2):280-290

Date

01/16/2020

Pubmed ID

31939843

DOI

10.1097/ALN.0000000000003061

Scopus ID

2-s2.0-85077864515 (requires institutional sign-in at Scopus site)   14 Citations

Abstract

BACKGROUND: Concern remains over reliable point-of-care testing to guide reversal of rivaroxaban, a commonly used factor Xa inhibitor, in high-acuity settings. Thromboelastography (TEG), a point-of-care viscoelastic assay, may have the ability to detect the anticoagulant effect of rivaroxaban. The authors ascertained the association of apparent rivaroxaban concentration with thromboelastography reaction time, i.e., time elapsed from blood sample placement in analyzer until beginning of clot formation, as measured using TEG and TEG6S instruments (Haemonetics Corporation, USA), hypothesizing that reaction time would correlate to degree of functional factor Xa impairment.

METHODS: The authors prospectively performed a diagnostic accuracy study comparing coagulation assays to apparent (i.e., indirectly assessed) rivaroxaban concentration in trauma patients with and without preinjury rivaroxaban presenting to a single center between April 2016 and July 2018. Blood samples at admission and after reversal or 24 h postadmission underwent TEG, TEG6S, thrombin generation assay, anti-factor Xa chromogenic assay, prothrombin time (PT), and ecarin chromogenic assay testing. The authors determined correlation of kaolin TEG, TEG6S, and prothrombin time to apparent rivaroxaban concentration. Receiver operating characteristic curve compared capacity to distinguish therapeutic rivaroxaban concentration (i.e., greater than or equal to 50 ng/ml) from nontherapeutic concentrations.

RESULTS: Eighty rivaroxaban patients were compared to 20 controls. Significant strong correlations existed between rivaroxaban concentration and TEG reaction time (ρ = 0.67; P < 0.001), TEG6S reaction time (ρ = 0.68; P < 0.001), and prothrombin time (ρ = 0.73; P < 0.001), however reaction time remained within the defined normal range for the assay. Rivaroxaban concentration demonstrated strong but not significant association with coagulation assays postreversal (n = 9; TEG reaction time ρ = 0.62; P = 0.101; TEG6S reaction time ρ = 0.57; P = 0.112) and small nonsignificant association for controls (TEG reaction time: ρ = -0.04; P = 0.845; TEG6S reaction time: ρ = -0.09; P = 0.667; PT-neoplastine: ρ = 0.19; P = 0.301). Rivaroxaban concentration (area under the curve, 0.91) and TEG6S reaction time (area under the curve, 0.84) best predicted therapeutic rivaroxaban concentration and exhibited similar receiver operating characteristic curves (P = 0.180).

CONCLUSIONS: Although TEG6S demonstrates significant strong correlation with rivaroxaban concentration, values within normal range limit clinical utility rendering rivaroxaban concentration the gold standard in measuring anticoagulant effect.

Author List

Myers SP, Dyer MR, Hassoune A, Brown JB, Sperry JL, Meyer MP, Rosengart MR, Neal MD

Author

Mitchell R. Dyer MD Assistant Professor in the Surgery department at Medical College of Wisconsin




MESH terms used to index this publication - Major topics in bold

Aged
Aged, 80 and over
Cohort Studies
Dose-Response Relationship, Drug
Factor Xa Inhibitors
Female
Humans
Male
Middle Aged
Point-of-Care Testing
Prospective Studies
Rivaroxaban
Thrombelastography